Year : 2017  |  Volume : 11  |  Issue : 2  |  Page : 79-80
Hemolysis during component preparation: An inadvertent cause


Department of Transfusion Medicine, PGIMER, Chandigarh, India

Click here for correspondence address and email

Date of Submission16-Jul-2016
Date of Acceptance20-Mar-2017
Date of Web Publication11-Sep-2017
 

How to cite this article:
Sachdeva P, Gupta A, Dhawan HK, Sharma RR, Marwaha N. Hemolysis during component preparation: An inadvertent cause. Asian J Transfus Sci 2017;11:79-80

How to cite this URL:
Sachdeva P, Gupta A, Dhawan HK, Sharma RR, Marwaha N. Hemolysis during component preparation: An inadvertent cause. Asian J Transfus Sci [serial online] 2017 [cited 2017 Sep 21];11:79-80. Available from: http://www.ajts.org/text.asp?2017/11/2/79/214353


We introduced a new type of blood bags from two different manufacturers at our center with diversion pouch and sampling device. Stripping of tube segments was also introduced to have anticoagulated blood in satellite tube. During component preparation, we observed 11 blood bags were showing pinkish to red colored plasma immediately after first centrifugation which was suggestive of hemolysis [Figure 1]. These bags were collected during outdoor blood donation camps and in-house collection on different occasions over 2-month period out of total collection of approximately 10,000.

Plasma hemoglobin in the bags was measured using Plasma/Low Hb System, HemoCue ®, Sweden, and percentage hemolysis was calculated using following formula:[1],[2],[3]
Figure 1: Normal supernatant plasma in first bag and pinkish to red discoloration of plasma due to hemolysis in next three bags after first centrifugation for component preparation

Click here to view




Plasma hemoglobin in these bags ranged from 2.8 to 4.4 g/L and percent hemolysis ranged from 1.2% to 2.18%. As per quality control criteria, hemolysis at the end of storage should be <1% in the United States or <0.8% in the European Union.[4] All these bags had percentage hemolysis of >1% on day of collection. We looked into the causes of hemolysis starting from blood collection to component preparation as shown in fishbone diagram [Figure 2]. Blood bags used were of standard specifications and ISO 3826 certified. As hemolysis was found in three bags in a single day, so donor-related factors could be ruled out. The G6PD levels were normal in all these bags. There were no phlebotomy-related problems (i.e., traumatic phlebotomy or hematoma) and collection time was within limits. Anticoagulant to blood ratio was maintained as there was no under collection. Bacterial contamination was ruled out as the culture from these bags was sterile. Component preparation was done using regularly calibrated refrigerated centrifuge. The most probable reason for hemolysis was narrowed down to shear stress due to turbulent flow caused by stripping across partially opened or closed Robert clamp as described earlier in literature.[1] Results were reproduced on quality control bags as deliberate stripping tube against closed clamp even once produces a similar amount of hemolysis in the bag. These packed red blood cell (PRBC) bags were used for this experiment just before discarding after routine quality control. Another clue to this cause, during investigation, it was found that in some of the bags, stripping was done by untrained helper staff during rush of donors in outdoor blood donation camps.
Figure 2: Fishbone diagram for causes of hemolysis during component preparation

Click here to view


To take corrective and preventive action, stripping is temporarily stopped in blood donation camps but is continued in in-house blood collection where all the staff nurses are under retraining by rotation. During 3-month follow-up, it was noted that out of 12,000 blood bags collected at camps, none of the bags showed hemolysis whereas similar hemolysis was observed in 1 out of 3000 bags collected during in-house collection where staff is under training.


   Conclusion Top


Hemolysis due to turbulent shear stress on RBCs can occur due to kinks in the tubes, partial opening of inline seal or partially opened entry ports at the time of stripping.[1],[3] This incident emphasizes that before the introduction of any new type of blood bags or new technique, proper training should be given to all staff members and compliance to standard operating procedure should be observed regularly.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Sowemimo-Coker SO. Red blood cell hemolysis during processing. Transfus Med Rev 2002;16:46-60.  Back to cited text no. 1
[PUBMED]    
2.
Makroo RN, Raina V, Bhatia A, Gupta R, Majid A, Thakur UK, et al. Evaluation of the red cell hemolysis in packed red cells during processing and storage. Asian J Transfus Sci 2011;5:15-7.  Back to cited text no. 2
[PUBMED]  [Full text]  
3.
Sawant RB, Jathar SK, Rajadhyaksha SB, Kadam PT. Red cell hemolysis during processing and storage. Asian J Transfus Sci 2007;1:47-51.  Back to cited text no. 3
[PUBMED]  [Full text]  
4.
Dumont LJ, Aronson CA, Dumont DF. Whole-blood collection and component processing. In: Fung MK, Grossman BJ, Hillyer CD, Westhoff CM, editors. Technical Manual. 18th ed. Bethesda, Maryland, USA: AABB; 2014. p. 135-66.  Back to cited text no. 4
    

Top
Correspondence Address:
Hari Krishan Dhawan
Department of Transfusion Medicine, PGIMER, Chandigarh - 160 012
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ajts.AJTS_89_16

Rights and Permissions


    Figures

  [Figure 1], [Figure 2]



 

Top
 
  Search

  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
    Email Alert *
    Add to My List *
* Registration required (free)  


   Conclusion
    References
    Article Figures

 Article Access Statistics
    Viewed119    
    Printed7    
    Emailed0    
    PDF Downloaded0    
    Comments [Add]    

Recommend this journal

Association Contact us | Sitemap | Advertise | What's New | Feedback | Copyright and Disclaimer

2006 - Asian Journal of Transfusion Science | Published by Wolters Kluwer - Medknow
Online since 10th November, 2006