Comparison of plateletpheresis on three continuous flow cell separators


Department of Transfusion Medicine, Tata Memorial Hospital, Mumbai - 400 012, India

Correspondence Address:
Anita Tendulkar
6th Floor, Service block, Department of Transfusion Medicine, Tata Memorial Hospital, E. Borges Rd., Parel, Mumbai - 400 012
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0973-6247.53877

Rights and Permissions

Introduction: Platelet concentrate (PC) remains one of the most important support measures in thrombocytopenic patients. An efficient cell separator is a prerequisite for an optimally functioning apheresis setup. Donor blood count may undergo a temporary reduction after the procedure. Aim: The aim was to find the extent of reduction in donor blood count (hemoglobin, hematocrit, white blood cell, and platelet) after plateletpheresis and to evaluate the cell separator for collection efficiency, processing time, and leukoreduction. Study Design and Methods: Two hundred and thirty seven procedures performed on the Amicus (N = 121), Fenwal CS-3000 Plus (N = 50) and Cobe spectra (N = 66) in a one year period were evaluated. The procedures performed on the continuous flow centrifugation (CFC) cell separators and donor blood counts (pre and post donation) done were included in the study. Results: The percent reduction in hemoglobin (HB), hematocrit (HCT), white blood cell (WBC) and platelet count ((PLT ct) was 2.9, 3.1, 9, 30.7 (Mean, N = 237) respectively after the procedure. The post donation PLT ct reduced to <100x109/L (range 80-100) in five donors (N = 5/237, Amicus). The pre donation PLT ct in them was 150-200x109/L. Collection efficiency (percent) of Amicus (79.3) was better as compared to the other two machines (CS: 62.5, Cobe: 57.5). PC collected on Cobe spectra had <1x106 WBC. The donor pre donation PLT levels had a positive correlation to the product PLT yield (r = 0.30, P = 0.000). Conclusion: Monitoring donor blood counts helps to avoid pheresis induced adverse events. A cautious approach is necessary in donors whose pre donation PLT ct is 150-200x109/L. The main variable in PLT yield is donor PLT ct (pre donation). High collection efficiency is a direct measure of an optimally functioning cell separator.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed5007    
    Printed268    
    Emailed2    
    PDF Downloaded224    
    Comments [Add]    
    Cited by others 10    

Recommend this journal

 

Association Contact us | Sitemap | Advertise | What's New | Feedback | Copyright and Disclaimer

2006 - Asian Journal of Transfusion Science | Published by Wolters Kluwer - Medknow
Online since 10th November, 2006